BIONIME Rightest GM300 Manual do Utilizador Página 2

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 2
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 1
- Healthcare professionals should evaluate their technique and their patients' technique at periodic
intervals. To accomplish this, it is recommended that BGM results be compared with a concurrently
obtained laboratory measurement on the same blood sample. A well characterized clinical
laboratory method employing hexokinase or glucose oxidase should be used as the comparative
method.
- Fluoride should not be used as a preservation for venous specimens when using blood glucose
monitors.
- Hands and fingers contaminated with sugar from foods or beverages may cause falsely elevated
results.
- Differences in whole blood and serum/plasma values may cause variability in results.
- Storage of strips near bleach as well as bleach containing products will affect results of glucose
oxidase strips.
- The use of cellular phones and other radio transmitting devices should be prohibited in areas where
testing occurs.
- Rightest™ Blood Glucose Test Strips are designed for use with capillary whole blood samples.
Do not use serum or plasma samples.
- Incorrect test results may be obtained at high altitude more than about 10000 feet (3048 meters)
above sea level.
- Hematocrits below 30% may cause higher results, and hematocrits above 55% may cause lower
results.
- Severe dehydration and excessive water loss may cause inaccurately low results.
- Rightest™ Blood Glucose Monitoring System has not been validated for use on neonates.
Therefore, it should not be used for neonates.
- Do not perform the blood glucose test at temperatures below 10 (50 ) or above 40 (104 ),
below 10% or above 90% relative humidity.
Storage and Handling
- Store the strips in the original capped vial at temperatures between 4 to 30 (39 to 86 ) and
relative humidity below 90%. Do not freeze.
- Replace the vial cap immediately and close tightly after taking test strip out from the vial. Do not
leave the cap of vial opened. If the strip is exposed in the air too long, it will absorb the moisture and
cause wrong test result.
- Use test strips within 3 months after first opening.
Measurement Range
The measurement range of the Rightest™ System is 10 to 600 mg/dL or 0.6 to 33.3 mmol/L.
Quality Control Section
Please refer to the Quality Control section of the User Manual.
Accuracy
The accuracy of the test study of the Rightest™ Blood Glucose Meter was demonstrated by comparing
whole blood (plasma equivalent) glucose values on the Rightest™ meter with plasma glucose values on a
lab instrument.
A total of 176 patients were enrolled. Each patient collected and tested their own blood samples (from the
fingertip) using the Rightest™ Meter. Another blood sample was collected within 5 minutes and got the
plasma to be analyzed on the lab instrument. Ninety-seven percent of the results were within +/- 20% of the
comparison method results at glucose concentrations 75 mg/dL and within +/-15 mg/dL at glucose
concentrations <75 mg/dL.
Troubleshooting and Customer Service
For more information on error messages and trouble shooting, please refer to the Error Messages
and Trouble Shooting section of the Rightest™ User manual.
You may also contact customer service by calling 886 4 24951268 ( 08:30 - 17:30, GMT + 08:00 ).
( At all other times, you could contact your healthcare professional for assistance )
Interferences
The following compounds may interfere with the glucose
measurement at the concentrations listed:
Uric acid > 0.54 mmol/L (> 9.0 mg/dL)
L -Dopa > 0.076 mmol/L (> 1.5 mg/dL)
Methyldopa > 0.071 mmol/L (> 1.5 mg/dL)
Cholesterol > 6.5 mmol/L (> 250 mg/dL)
Reagents
Each Blood Glucose Test Strip contains the following
reagents:
Glucose Oxidase (Aspergillus niger) (GOD) 8.5 %
Potassium ferricyanide 48.5 %
Non-reactive ingredients 43 %
References
1) Consensus statement on self monitoring of blood glucose. Diabetes Care 1987; 1:95-99.
2) Diabetes Information - American Diabetes Association and National Institute of Diabetes and
Digestive and Kidney Diseases ( NIDDK )
Store between tempe rature 4 and 30 ( 39 and 86 )
For in vitro diagnostic use
Manufacturer
LOT
IVD
Lot number
Use by
101-3GS300-034
EN
BIONIME CORPORATION
694, Renhua Road, Dali City,
Taichung County, Taiwan 412
Tel: +886 4 24951268 Fax: +886 4 24952568
E-mail: [email protected] http://www.bionime.com
EU Representative:
BIONIME GmbH
Heinrich Wild Strasse 202,
CH-9435 Heerbrugg
Switzerland
Accuracy
The accuracy of the test study of the Blood Glucose Meter was demonstrated by comparing
whole blood (plasma equivalent) glucose values on the Rightest™ meter with plasma glucose values on
a lab instrument.
A total of 176 patients were enrolled. Each patient collected and tested their own blood samples ( from
the fingertip ) using the Rightest™ Meter. Another blood sample was collected within 5 minutes and got
the plasma to be analyzed on the lab instrument.
Linearity regression produced the following line equation (Lab instrument-YSI 2300D was considered to
be the reference method):
Rightest™
Patient
Rightest fingertip vs YSI-Plasma
Test range
Test number
Slope
Intercept (mg/dL)
r
30~572 mg/dL (1.7~31.8 mmol/L)
176
1.03
-0.25
0.9911
0
50
100
150
200
250
300
350
400
450
500
550
600
0 50 100 150 200 250 300 350 400 450 500 550 600
N=176
Slope=1.03
Intercept=-0.25
r=0.9911
YSI (mg/dL)
Patient
Fingerstick (mg/dL)
E
D
B
B
D
C
C
E
A
A
Additional Information for Healthcare Professionals
Detection Principle
The glucose oxidase and potassium ferrocyanide in the strip react with the glucose in the sample to
produce an electrical current which is proportional to the amount of glucose in the sample. The meter
measures the current and converts it to the corresponding glucose concentration.
Performance Characteristics
Measurement Range
The measurement range of the Rightest™ System is 10 to 600 mg/dL or 0.6 to 33.3 mmol/L.
Precision
The precision was evaluated including (i) venous whole blood sample (ii) 3 levels glucose control
solution in period of 20 days, by 10 meters and 1 batch of strip.
(ii) Control solution:
P-01
100
40.6(2.3)
1.5(0.08)
3.6%
P-02
100
96.9(5.4)
2.5(0.14)
2.6%
P-03
100
117.3(6.5)
2.5(0.14)
2.1%
P-04
100
192.9(10.7)
3.9(0.22)
2.0%
P-05
100
351.5(19.5)
6.3(0.35)
1.8%
CS-L
100
50.5(2.3)
1.2(0.06)
2.5%
CS-N
100
102.2(5.8)
2.0(0.11)
1.9%
CS-H
100
302.3(16.7)
6.6(0.36)
2.2%
(i) Venous whole blood sample:
Glucose levels
(1) Total test numbers (n)
(2) Mean mg/dL (mmol/ L)
(3) SD mg/dL (mmol/ L)
(4) CV (%)
Meters
(1) Total test numbers (n)
(2) Mean mg/dL (mmol/ L)
(3) SD mg/dL (mmol/ L)
(4) CV (%)
Vista de página 1
1 2

Comentários a estes Manuais

ChrisTom 17 Apr 2024 | 22:41:43

Hello, im noob :)