Performing a Quality Control Test
Performing a Quality Control Test
8. When you hear a sound of " beep " , leave the meter on the table while waiting the
test result. The screen will show the countdown mode.
9. Tightly recap the cap of control solution.
10. The control result appears after the measurement is completed. Compare your
control test result to the range printed on strip vial label.
Your result should fall within the control solution range
printed on the label of strip vial. If result is not within the
control range, please refer to Understanding Control Test
Results on page 56.
Please follow the local regulation and discard the used strip
properly.
Example
Control Solution Range
90-130 mg/dL
5.0-7.2 mmol/L
PRECAUTION
- Your quality control test result by control solution test will not be calculated for average reading but still can be
recalled. The control solution test result will be shown with " " symbol on the screen.
- You have to do check key test before control solution test. If you only use the control solution to do the quality
control test, your test result will be memorized and calculated for average same as real blood test.
- Please don't touch the control solution by approaching the vial to the sample entry on
strip ( see drawing ). The reagent might be sucked into the vial of control solution which
alteration or degeneration might happen. If you approach the control solution vial to the
sample entry on strip for 3~5 times, you will find that the measurements might lower
down for 10-20%.
- Keep your Smart Code Key dry and clean. Please clean it up right away if your Smart
Code Key gets stain or moisture.
( Refer to page 57 for cleaning )
×